Vaisala HMP155 User Manual Page 65

  • Download
  • Add to my manuals
  • Print
  • Page
    / 84
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 64
Chapter 6 ___________________________________________________Calibration and Adjustment
VAISALA _______________________________________________________________________ 63
Serial Line Calibration
Two-Point Humidity Adjustment
Note that the difference between the two humidity references must be at
least 30 %RH.
To make a humidity adjustment:
1. Connect the HMP155 to a PC. See section Serial Line
Communication on page 24. Open a terminal program.
2. Carry out the chem
ical purge (if available). See section Chemical
Purge (Optional) on page 45.
3. Press the
ADJ button and the green LED lights up.
4. Remove the filter from the probe and insert the probe into a
measurement hole of the dry end reference chamber (for example,
LiCl: 11 % RH).
5. Enter the
CRH command and press ENTER.
6. Wait at least 30 minutes for the sensor to stabilize.
7. Press
ENTER a few times to check if the reading has stabilized.
8. When the reading has stabilized, give the reference humidity after
the question mark and press
ENTER.
>crh
RH : 16.6675 1. ref ?
RH : 16.4978 1. ref ?
RH : 16.3956 1. ref ? 11.25
Press any key when ready ...
RH : 11.25 Ref2 ?
9. Now the device is waiting for the high end reference. Insert the
probe into the measurement hole of the high end reference chamber
(for example, NaCl: 75 % RH chamber in the humidity calibrator
HMK15). Press any key when ready.
10. Let the probe stabilize for about 30 minutes. You can follow the
stabilization by pressing
ENTER.
11. When stabilized, type the high end reference value after the
question mark and press
ENTER.
RH : 75.45 Ref2 ?
RH : 75.57 Ref2 ?
RH : 75.55 Ref2 ?
RH : 75.59 Ref2 ? 75.5
OK
12. OK indicates that the adjustment has succeeded and the new
calibration coefficients are calculated and stored.
Page view 64
1 2 ... 60 61 62 63 64 65 66 67 68 69 70 ... 83 84

Comments to this Manuals

No comments