Vaisala HMW90 User Manual Page 78

  • Download
  • Add to my manuals
  • Print
  • Page
    / 110
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 77
User's Guide _______________________________________________________________________
76 ___________________________________________________________________ M211399EN-F
Error Messages
Error Messages on the Display
Table 12 Error Messages on the Display
Error Message
Possible Cause and Solution
HTM10 01
HTM10 02
Communication failure with HTM10
module. Reconnect the module and check that it sits
firmly in place.
HTM10 03
HTM10 04
HTM10 05
HTM10 06
Problem with the HTM10 module.
1. Check the module for damage.
2. Check if the humidity sensor is missing or loose. Note
that the humidity sensor is not included on TMW-type
transmitters since they only measure temperature.
3. Replace the module if unable to remove the problem.
Internal 1
Internal 2
Internal 3
Internal problem with the transmitter.
1. Reset the transmitter.
2. Restore the factory settings using service port or DIP
switches if reset does not help.
Error Messages on the Serial Line
View Currently Active Errors
Use the ERRS command to view currently active errors on the serial
line:
ERRS<cr>
Example:
>errs
NO ERRORS
Page view 77
1 2 ... 73 74 75 76 77 78 79 80 81 82 83 ... 109 110

Comments to this Manuals

No comments