Page view 37
USER'S GUIDE____________________________________________________________________
Environment Menu
Pressure Settings
In pressurized environments, the actual process pressure value must be
set for DM70. The pressure can be given in the following units:
P
barg:
Gauge pressure is given in the unit of bar. It
indicates the pressure difference between the normal
atmospheric pressure and the actual process pressure.
P
bara:
Absolute process pressure is given in the unit of bar.
P
psig:
Gauge pressure is given in the unit of psi. It indicates
the pressure difference between the normal
atmospheric pressure and the actual process pressure.
P
psia:
Absolute pressure is given in the unit of psi.
To set the values, follow the instructions below:
1. Open the menu by pressing Open.
2. Select Environment with and then press . See Figure
13 below.
0403-036
Figure 13 Environment Menu
3. To change the pressure unit, press Unit. The default unit is
barg.
4. To set the ambient pressure value, press Set.
5. Set the value using the buttons. To change the sign of
the pressure value, press
+/-. To save the value, press
OK.
6. To return to the basic display, press
Exit.