HMI41 UND SONDEN
Bedienungsanleitung U171de-1.3
48
Genauigkeit der absoluten
Feuchte (g/m
3
)
rF/%
10 20 30 40 50 60 70 80 90 100
T/°C
-40 0.003 0.003 0.003 0.003 0.004 0.004 0.004 0.005 - -
-20 0.020 0.021 0.023 0.025 0.026 0.028 0.029 0.031 - -
0 0.10 0.11 0.12 0.13 0.13 0.14 0.15 0.15 0.16 0.17
20 0.37 0.39 0.41 0.43 0.45 0.47 0.49 0.51 0.53 0.55
40 1.08 1.13 1.18 1.24 1.29 1.34 1.39 1.44 1.49 1.54
60 2.73 2.84 2.95 3.07 3.18 3.29 3.40 3.52 3.63 3.74
80 6.08 6.30 6.51 6.73 6.95 7.17 7.39 7.61 7.83 8.05
100 12.2 12.6 13.0 13.4 13.8 14.2 14.6 15.0 15.3 15.7
120 22.6 23.3 23.9 24.6 25.2 25.8 26.5 27.1 27.8 28.4
140 39.1 40.0 41.0 42.0 43.0 44.0 45.0 45.9 46.9 47.9
160 63.5 64.9 66.4 67.8 69.2 70.7 72.1 73.5 75.0 76.4
Genauigkeit der Feuchttemperatur
(°C)
rF/%
10 20 30 40 50 60 70 80 90 100
T/°C
-40 0.20 0.20 0.20 0.20 0.20 0.20 0.20 0.20 - -
-20 0.21 0.21 0.21 0.21 0.22 0.22 0.22 0.22 - -
0 0.27 0.28 0.28 0.29 0.29 0.29 0.30 0.30 0.31 0.31
20 0.45 0.45 0.45 0.44 0.44 0.44 0.43 0.43 0.42 0.42
40 0.84 0.77 0.72 0.67 0.64 0.61 0.58 0.56 0.54 0.52
60 1.45 1.20 1.03 0.91 0.83 0.76 0.71 0.67 0.63 0.61
80 2.24 1.64 1.32 1.13 0.99 0.90 0.82 0.76 0.72 0.68
100 3.06 2.04 1.58 1.31 1.14 1.01 0.92 0.85 0.80 0.75
120 3.86 2.41 1.81 1.48 1.28 1.13 1.03 0.95 0.88 0.83
140 4.57 2.73 2.03 1.65 1.41 1.25 1.13 1.04 0.97 0.91
160 5.23 3.04 2.24 1.81 1.55 1.36 1.23 1.13 1.05 0.98
11.6. Elektromagnetische Verträglichkeit
11.6.1. Störaussendung
Gestrahlte Störgrößen, Test nach EN55022
11.6.2. Störfestigkeit
Test: Test nach: Leistung:
Gestrahlte Störgrößen IEC 1000-4-3 Klasse 3
Schnelle transiente elektrische
Störgrößen IEC 801-4 Klasse 4
Elektrostatische Entladung IEC 801-2
Page view 51