User's Guide _______________________________________________________________________
28 ___________________________________________________________________ M211244EN-B
Set/Show Available Quantities
The CALCS command shows the abbreviations of the quantities
measured by HMT120. The command can also be used to select the
desired quantities for measurement. Only after the quantities have been
set with the CALCS command, can these quantities be selected for
different outputs (see commands DSEL, ASEL, and FORM). The
quantities in DSEL, ASEL, and FORM must match with CALCS.
CALCS [q1 q2]<cr>
where
q1, q2
=
Any two of the allowed quantity parameters, i.e. RH, T,
TD, TDF, A, X, H, TW, PWS, PW
Examples:
>calcs
RH Td
>
calcs td t
>
NOTE
Single-parameter transmitters HMT120H and HMT120T have only one
quantity (T for HMT120T, and RH, TD, TDF, A, X, H, TW, PWS, PW
for HMT120H.)
Set/Show Calibration Date
Use the CDATE command to show or set the calibration date field.
(Calibration refers to the calibration of the HMT120's analog outputs.)
CDATE [date]<cr>
where
date
=
Calibration date in format yyyy-mm-dd
Examples:
>cdate
Calibrated : "NOT SET"
>
>cdate 2009-06-22
Calibrated : "2009-06-22"
>