Appendix B ______________________________________________________ Calculation Formulas
VAISALA ______________________________________________________________________ 155
The water vapor pressure is calculated using:
100
ws
w
P
RHP
(7)
Parts per million by volume is calculated using:

w
w
v
Pp
P
ppm
6
10
(8)
Symbols:
T
d
= dewpoint temperature (°C)
P
w
= water vapour pressure (hPa)
P
ws
= water vapour saturation pressure (Pa)
RH
= relative humidity (%)
x
= mixing ratio (g/kg)
p
= atmospheric pressure (hPa)
a
= absolute humidity (g/m
3
)
T
= temperature (K)
h
= enthalpy (kJ/kg)