Other Commands
Check Transmitter Settings
Syntax: ?<cr>
Syntax: ??<cr>
Use the ? command to check the current transmitter configuration. The
?? command is similar but can also be used if the transmitter is in POLL
mode.
Example (factory default settings):
>?
MMT310 / 1.03
PRB serial nr : V1234567
Calibration : 2003-03-25
Cal. info : NU/HMK15
Output units : metric
Pressure : 1013.25 hPa
RS232 settings
Address : 2
Output interval: 1 MIN
Baud P D S : 4800 E 7 1
Serial mode : STOP
Analog outputs
Ch1 output mode: 0 ...20mA
Ch2 output mode: 4 ...20mA
Ch1 error out : 0.000mA
Ch2 error out : 0.000mA
Ch1 T lo : -40.00 'C
Ch1 T hi : 60.00 'C
Ch2 ppm lo : 0.00 ppm
Ch2 ppm hi : 5000.00 ppm
>