Table 20 Temperature Specifications and Operating Pressure
Ranges
Property
Description / Value
HMT311
-40 ... +60 °C (-40 ... +140 °F)
HMT313 80 °C
-40 ... +80 °C (-40 ... +176 °F)
HMT313 120 °C
-40 ... +120 °C (-40 ... +248 °F)
HMT314
-70 ... +180 °C (-94 ... +356 °F),
0 ... 10 MPa (0 ... 100 bar)
HMT315 (vapor tight)
-70 ... +180 °C (-94 ... +356 °F)
HMT317 (vapor tight)
-70 ... +180 °C (-94 ... +356 °F)
HMT318
-70 ... +180 °C (-94 ... +356 °F),
0 ... 4 MPa (0 ... 40 bar)
Accuracy at +20 °C (+68 °F)
±0.2 °C (±0.36 ˚F)
Accuracy over temperature range
See Figure 19 below.
Temperature sensor
Pt 100 RTD Class F0.1 IEC 60751
0507-021
Figure 19 Accuracy over Temperature Range