Vaisala DPT145 User's Guide Page 54

  • Download
  • Add to my manuals
  • Print
  • Page
    / 63
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 53
User's Guide _______________________________________________________________________
54 ___________________________________________________________________ M211371EN-D
CHAPTER 8
TECHNICAL DATA
This chapter provides the technical data of the DPT145.
Specifications
Table 12 Measured Parameters
Parameter
Measured Range
Dewpoint
-50 ... +30 °C (-58 ... +86 °F)
Pressure, absolute
1 ... 12 bar (14.5 ... 174 psi)
Temperature
-40 ... +80 °C (-40 ... +176 °F)
Table 13 Calculated Parameters
Parameter
Measured Range
Pressure, normalized to 20 °C (68 °F)
1 ... 12 bara (14.5…174 psia)
SF
6
or SF
6
/N
2
mixture density
0 ... 100 kg/m
3
ppm moisture, by volume
40 ... 40 000 ppm
Dewpoint, converted to atmospheric
pressure
-65 … +30 °C (-85 ... +86 °F)
Table 14 Performance
Property
Description / Value
Dewpoint accuracy
±3 °C (±5.4 °F), see Figure 12 on
page 55
Dewpoint stability
typical drift < 2 °C (3.6 °F)/5a
Pressure accuracy at 23 °C (73.4 °F)
±0.04 %FS
Pressure temperature dependence
±0.01 bar/10 °C (18 °F)
Temperature accuracy
0 ... +40 °C (+32 ... +104 °F)
±0.5 °C (± 0.9 °F)
-40 ... +80 °C (-40 ... +176 °F)
±1 °C (± 1.8 °F)
Density accuracy (pure SF
6
, 1 ... 10 bara)
0 ... 40 °C (+32 ... +104 °F)
±1 %FS
-40 ... +60 °C (-40 ... +140 °F)
±2.2 %
ppm accuracy, typical
(5 ... 1000 ppm, 7 bar)
±(7 ppm + 15 % of reading)
Sensor response time
Pressure response time
< 1 s
Dewpoint response time*
63% [90%] at 20°C and 1 bar
-50 -> -10 °C Tdf
5 s [10 s]
-10 -> -50 °C Tdf
10 s [2.5 min]
* system equilibrium related response time is typically longer
Page view 53
1 2 ... 49 50 51 52 53 54 55 56 57 58 59 ... 62 63

Comments to this Manuals

No comments