Vaisala DPT145 User's Guide Page 55

  • Download
  • Add to my manuals
  • Print
  • Page
    / 63
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 54
Chapter 8 _____________________________________________________________ Technical Data
VAISALA ________________________________________________________________________ 55
1107-032
Figure 12 DPT145 Dewpoint Measurement Accuracy
Table 15 Operating Environment
Property
Description / Value
Operating temperature of electronics
-40 … +60 °C (-40 ... +140 °F)
Pressure
0 … 50 bar (0…725 psi)
Relative humidity
0 ... 100 %
Measured gases
SF
6
SF
6
/N
2
mixture
Table 16 Outputs
Property
Description / Value
Digital output
RS-485, non-isolated, Vaisala protocol
Connector
4-pin M8
0 20
20
0
10
30
-
10
-
20
-
3
0
10 30
4
0 5
0 6
0
4
0
Dew
po
int
te
mpera
ture (°
C)
Temperature of meas
u
re
d
ga
s (°C)
-40 -30 -20 -10 7
0 8
0
-4
0
-50
-60
Not recommended for
c
ontinuous measurement
Dewpoint measurement low limit
-70
Accurac
y
no
t s
p
ec
if
i
ed
Accurac
y
±
3 °
C
Td