USER'S GUIDE____________________________________________________________________
68 _______________________________________________________________ M210912EN-C
ACAL [0/1]
Example:
Ch 0:
>acal 0
U1 ( V ) ? 1.001
U2 ( V ) ? 9.011
Ch 1:
>acal 1
U1 ( V ) ? 2.0988
U2 ( V ) ? 8.8997
>
MI70 Check and Adjustment
Vaisala MI70 measurement indicator is an optional accessory that can be
used as a display or as a communication device for HMP155. When it is
used, HMP155 is powered via MI70.
With MI70 you can check the HMP155 easily on the field. Both the
active output version and the passive output version (only RH displayed)
can be checked with MI70.
In the active output version relative humidity, temperature and additional
T-probe temperature can be calibrated. In the passive output version,
only relative humidity calibration is possible.
To check and adjust HMP155 with MI70:
1. Connect HMP155 to MI70 by using a connection cable (Vaisala
item 221801).
2. Turn MI70 on by pressing the power key and follow the
instructions given on the display.
NOTE
When turning on the passive output version, always hold down the ADJ
button simultaneously.
3. Activate the adjustment mode by holding down the
ADJ button on
the probe until the message below appears on the MI70 display.