Chapter 3 _______________________________________________________ AWS Client Software
VAISALA_______________________________________________________________________ 87
The log entry information includes the time tag, that is, date and time of
the entry, the status, and the value of the logged measurement or
calculation.
Controlling Logging
Logging is automatically on if it has been defined in the setup and if it
has not been stopped. Logging is automatically stopped when the log
file is retrieved with AWS Client software. When the download is
completed, logging is activated again automatically.
When a sensor is replaced, stopping is not necessary if some invalid log
items can be tolerated. Alternatively, instead of stopping logging, you
can manually disable a measurement or a sensor. For detailed
information, see section Measurement Enable or Disable on page 88.
To see the logging groups, type logstatus. To see the current logging
status of a certain group, type logstatus <group_id>. To stop or start the
logging of a certain group, type logstop/loggo <group_id>. For the
output of the commands, see Figure 47 on page 88.
Table 16 Log Entry Status
Status Indicator Status Description
-I----- Invalid Value may be outside the set
scale; that is, the set
climatological limits or step
change validation.
-I--N- Invalid; not available No measurements done yet
V----- Valid (normal) Measurement or calculated
value available normally