User's Guide _______________________________________________________________________
76 ___________________________________________________________________ M210762EN-A
5. Select RUN interval for RUN communication mode and press
SET to confirm your selection.
6. Use the arrow buttons to set the measuring interval and the unit.
Press OK to confirm your settings.
7. Select POLL address for POLL communication mode. Press
SET to confirm your selection.
8. Use the arrow buttons to set the transmitter address. Press OK to
confirm the setting.
9. Use the arrow buttons to select ECHO. Press ON to turn to it
on. Press OFF to turn it off.
10. Press EXIT to return to the basic display.
SERI
Use the serial line command SERI [b p d s] to set communication
settings for the user port.
Syntax: SERI [b p d s]
where
b = Bauds (110, 150, 300, 600, 1200, 2400, 4800, 9600,19200,
38400, 57600, 115200)
p = Parity (n = none, e = even, o = odd)
d = Data bits (7 or 8)
s = Stop bits (1 or 2)
Example:
>SERI 600 N 8 1
600 N 8 1
>
SMODE
Use the command SMODE [xxxx] to set the user port start-up
operating mode.
Syntax: SMODE [xxxx]
where
xxxx = STOP, RUN, POLL or SEND