Vaisala DRYCAP DMT346 User's Guide Page 88

  • Download
  • Add to my manuals
  • Print
  • Page
    / 146
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 87
User's Guide _______________________________________________________________________
86 ___________________________________________________________________ M210762EN-A
1. Select the current/voltage output, switch ON either of the
switches, 1 or 2.
2. Select the range, switch ON one of the switches from 3 to 7.
0503-045
Figure 45 Current/Voltage Switches of Output Modules
Numbers refer to Figure 45 above:
1 = Current/voltage selection output switches (from 1 to 2)
2 = Current/voltage range selection switches (from 3 to 7) in
analog output 1 and 2.
3 = Switches for service use only. Keep in OFF position always.
NOTE
Only one of the switches, 1 or 2, must be ON at a time.
Only one of the switches, 3 to 7, must be ON at a time.
Page view 87
1 2 ... 83 84 85 86 87 88 89 90 91 92 93 ... 145 146

Comments to this Manuals

No comments