Vaisala HUMICAP MMT310 User's Guide Page 38

  • Download
  • Add to my manuals
  • Print
  • Page
    / 63
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 37
USER'S GUIDE____________________________________________________________________
36 __________________________________________________________________ M210474EN-B
Low end adjustement
1.
Remove the transmitter unit from the mounting plate (see page
12), and press the adjustment button once, see picture.
2. Remove the filter from the probe and insert the probe head into
a measurement hole of the dry end reference chamber (e.g. LiCl:
11 % RH in the humidity calibrator HMK15, please use the
adapter fitting (13.5 mm)).
3. Wait at least 30 minutes for the sensor to stabilize.
4. Give command CRH and press ENTER.
CRH
Type C and press ENTER a few times to check if the reading is
stabilized.
5. When the reading is stabilized, give the reference humidity after
the question mark and press ENTER.
>crh
RH : 11.25 Ref1 ? c
RH : 11.25 Ref1 ? c
RH : 11.25 Ref1 ? c
RH : 11.24 Ref1 ? c
RH : 11.24 Ref1 ? 11.3
Press any key when ready ...
Now the device is waiting for the high end reference.
Adjustment button
Page view 37
1 2 ... 33 34 35 36 37 38 39 40 41 42 43 ... 62 63

Comments to this Manuals

No comments