Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 39
LI Reverting the factory calibration
1. Remove the transmitter unit from the mounting plate (see page 12)
and press the adjustment button, see page 26.
2. Type command LI and give value 0 for an offset value and value 1
for a gain value.
LI
>li
RH offset : -0.6000000 ? 0
RH gain : 1.00000000 ? 1
T offset : 0.00000000 ? 0
T gain : 0.40000000 ? 1
>
3. Reset the transmitter by giving a command RESET. Transmitter
returns to normal mode
FCRH Relative humidity calibration after sensor change
FCRH
The transmitter asks and measures relative humidity and calculates the
calibration coefficients. This 2-point adjustment should be performed
after a sensor change. Follow the more detailed calibration
instructions on page 35, but instead of giving CRH command, use
FRCH command.
>FCRH
RH : 1.82 1. ref ? 0
Press any key when ready...
RH : 74.22 2. ref ? 75
OK
>
The OK indicates that the calibration has succeeded.
CTEXT Giving text to calibration information field
CTEXT