Chapter 5 ___________________________________________________Calibration and Adjustment
VAISALA _______________________________________________________________________ 23
1. Choose the COM port in use from the file menu.
2. To open the connection click the Read PTB210-button.
3. Give reference pressure values and the corresponding corrections.
Calculate the corrections according to the following equation. The
settings on left-hand side can also be changed.
measuredreference
PPcorr
(1)