Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 99
Getting the Measurement Message from
Serial Line
Starting Continuous Outputting
Enter the R command to start the continuous output of measurements.
R<cr>
Example:
>r
Tdf=-20.6 'C H2O= 958 ppmV x= 0.6 g/kg
>r
Tdf=-20.7 'C H2O= 958 ppmV T= 23.8 'C RH= 3.3 %RH
If a value is too long to fit to the allocated space in the output, or if there
is an error in outputting the quantity, the value is displayed with stars ‘*’.
You can change the format of the output with the following commands:
- Outputting interval can be changed with the INTV command.
- Output message format can be changed with the FORM command.
Stopping Continuous Outputting
Use the S command to end the RUN mode. After this command all other
commands can be used. You can also press the Esc button or reset the
transmitter to stop the outputting.
S<cr>
See command SMODE to change the default (power-up) operation
mode.