User's Guide _______________________________________________________________________
44 ___________________________________________________________________ M210704EN-G
Ball Valve Installation Kit for DMT348
The ball valve installation kit (Vaisala order code: BALLVALVE-1) is
preferred when connecting the probe to a pressurized process or pipeline.
Use the ball valve set or a 1/2" ball valve assembly with a ball hole of
ø14 mm or more. If you install the probe (ø12 mm) in a process pipe,
please note that the nominal size of the pipe must be at least 1 inch (2.54
cm). Use the manual press handle to press the probe into the pressurized
(< 10 bar) process or pipeline.
0507-043
Figure 30 Installing the DMT348 Probe Through a Ball Valve
Assembly
The numbers refer to Figure 30 above:
1
=
Manual press tool
2
=
Handle of the ball valve
3
=
Probe
4
=
Process chamber or pipeline
5
=
Groove on the probe indicates the upper adjustment limit
6
=
Filter
7
=
Ball of the ball valve
8
=
Fitting screw