APPENDIX D ____________________________________________________ PULSE OUTPUT MODE
VAISALA _________________________________________________________________________ 83
APPENDIX D PULSE OUTPUT MODE
NOTE
The pulse output mode can be used with hPa/mbar pressure units only.
The pulse output mode of the PTB220 series barometers is useful
when no serial interface is available in the host system. A standard
pulse counter interface is all that is needed to count the number of
pulses output by the barometer. The host system must create a trigger
pulse for the barometer, and reset and open the gate of the counter to
be able to receive the pulses. The pulse counter must be capable of
measuring pulse rates up to 5 kHz or 50 kHz.
The trigger pulse must be a TTL level signal. The pulse output of the
PTB220 series barometers is triggered at the falling edge of the trigger
pulse. The pulses are then output within 1 to 100 ms depending on the
internal status and settings of the barometer. The recommended
procedure is to first reset and open the gate of the counter and then to
give the trigger pulse to the barometer. The gate-open time must be
long enough to allow enough time for all pulses to be output by the
barometer. The pulse output and trigger input signal levels are at
5VDC level when the barometer is outputting no pulses. The trigger
pulse can be a positive pulse or negative pulse but its duration must be
minimum 1 µs.
The figure below illustrates the timing principle of pulse output mode:
output pulses
counter gate open
trigger pulse min. 1 us, trigger at falling edge of the pulse
pulse output starts within 1...100 ms from trigger