Chapter 4 ________________________________________________________________ Operation
VAISALA_______________________________________________________________________ 39
SETDIR average <value>
Example:
>setdir average 3.00
AVERAGE=3.00
>
SETMES
The SETMES command specifies the data transmission mode. The
transmitter can support four (4) different messages at a time. Each
message has its own settings for the message type and transmit interval.
Three message types are supported:
- NONE (no message)
- MWV (standard wind speed and direction message)
- MWVQUERY (Vaisala extension to MWV message)
SETMES {messagenumber} [type] [interval] <value><cr>
where
value = Wind direction computation time period in seconds. Range
0.25 ... 5.00 in 0.25 second increments (default 3.00)
where
messagenumber = The number of the message to configure. Range 0
... 3
type = Configures the message type. Possible values:
NONE
MWV
MWVQUERY
interval = Configures the transmit interval. Zero value
disables data auto-transmission. Possible values: 0
... 600.00 in 0.25 second increments
value = The value for the option.