Models
Model number Type Output Special Features Ingress Protection
HMW110 Wall-mount, RH+T 2-wire, current output Configurable model* IP65
HMW112 Wall-mount, RH+T 2-wire, current output IP65
HMD110 Duct-mount, RH+T 2-wire, current output Configurable model* IP65
HMD112 Duct-mount, RH+T 2-wire, current output IP65
HMS110 Outdoor, RH+T 2-wire, current output Radiation shield, configurable model* IP65
HMS112 Outdoor, RH+T 2-wire, current output Radiation shield IP65
*Delivered with customer specific output settings, including calculated humidity parameters and special scaling of outputs.
Technical Data
Performance
RELATIVE HUMIDITY
Measurement range 0 ... 100 %RH
Accuracy
Temperature range +10 ... +30 °C (+50 ... +86 °F)
0 ... 90 %RH ±2 %RH
90 ... 100 %RH ±3 %RH
Temperature range -20 ... +10 °C, +30 ... +60 °C
(-4 ... +50 °F, +86 ... +140 °F)
0 … 90 %RH ±3 %RH
90 … 100 %RH ±4 %RH
Temperature range -40 ... -20 °C (-40 ... -4 °F)
0 ... 100%RH ±4 %RH
Stability in typical HVAC applications ±0.5 %RH/year
Humidity sensor Vaisala HUMICAP
®
180R
TEMPERATURE
Measurement range -40 ... +60 °C (-40 ... +140 °F)
Accuracy
At +20 °C (+68 °F) ±0.3 °C
Temperature dependence ±0.01 °C/°C
Temperature sensor Pt1000 RTD Class F0.1 IEC 60751
CALCULATED PARAMETERS
Measurement range for dew
point temperature and wet bulb
temperature -40 ... +60 °C (-40 ... +140 °F)
Measurement range for
enthalpy -40 ... 460 kJ/kg (-10 ... +190 BTU/lb)
Accuracy of the calculated parameters should be
calculated at the actual condition based on the RH
and temperature specification.
Accuracy at 20°C (68°F) and 80%RH:
Dew point ±0.7 °C (1.2 °F)
Wet bulb temperature ±0.5 °C (0.9 °F)
Enthalpy ±1.6kJ/kg (0.7 BTU/lb)
Operating Environment (All Models)
Operating temperature range -40 ... +60 °C (-40 ... +140 °F)
Operating humidity range 0 ... 100 %RH
Maximum wind/flow speed 30 m/s
Storage temperature -40 ... +60 °C (-40 ... +140 °F)
Electromagnetic compliance EN61326-1, Industrial Environment
Mechanics
Max wire size 1.5 mm
2
(AWG 16)
Standard housing color White (RAL9003)
Housing material PC + 10%GF (UL-V0 approved)
Inputs and Outputs
Analog output 4 ... 20 mA, loop powered
Loop resistance 0 ... 600 Ω
Supply voltage 20 ... 28VDC at 600 Ω load
10 ... 28VDC at 0 Ω load
Data input for RDP100 Remote
Panel Display
RS485,
Vaisala proprietary protocol
Spare Parts and Accessories
Remote Panel Display RDP100
Conduit fitting + O-ring (M16x1.5 / NPT1/2 Inch) 210675SP
Conduit fitting + O-ring (M16x1.5 / PG9, RE-MS) 210674SP
Fastening set HMS110 237805
Porous PTFE Filter DRW239993SP
Membrane Filter ASM210856SP
Terminal Block, Blue 236620SP
USB cable for PC connection 219690
Connection cable for HM70 hand-held meter 219980SP
HUMICAP
®
180R sensor HUMICAP180R
Catalytic HUMICAP
®
sensor HUMICAP180VHP
Ref. B211349EN-B ©Vaisala 2014
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
Please contact us at
www.vaisala.com/requestinfo
www.vaisala.com
Scan the code for
more information
Page view 11
1 2 ... 7 8 9 10 11 12 13 14 15 16 17 ... 27 28

Comments to this Manuals

No comments