CHAPTER 4_____________________________________________________________COMMISSIONING
VAISALA _________________________________________________________________________ 21
Example of setting the limit to 0.5 hPa:
>pdmax <cr>
Pd max : 1.000 ? 0.5 <cr>
>
Example of exceeding the Pdmax limit:
>form <cr>
4.2P""UUUU \r \n
?4.2P1""P2""P""UUU""ERR\r\n<cr>
>send <cr>
1020.30 1020.32 1020.31 hPa 00
>send <cr>
1020.30 1022.30 1021.30 hPa 11
>
Please use the ERRS command to analyze problems.
KEYLOCK Setting the keyboard lock on/off
KEYLOCK x <cr>
where:
x = ON or OFF
The KEYLOCK command is used to set or inspect the keypad lock
condition of the display cover keypad. The PTU200 transmitters with
display cover are supplied with the keypad locked (KEYLOCK OFF).
It is thus possible to inspect the settings but not to change them. With
the keypad locked the settings can be inspected but not changed.
Example:
>keylock on <cr>
Keylock : ON
>
If anyone tries to change the settings using the keypad with the
KEYLOCK ON, the transmitter will display NO MODIFICATIONS
ALLOWED
for a few seconds and then returns to the original display.
Hardware settings
Inside the transmitter, there is a connector X15 (see FIGURE 4-2) and a
dip switch S1 (see
FIGURE 4-3). These are used to make the main
hardware settings of the PTU200.