User's Guide ______________________________________________________________________
114 _________________________________________________________________ M210796EN-A
0508-012
Figure 46 Chemical Purge Settings
- Select Automatic purge and turn it on or off by pressing the
ON/OFF key.
- Select Interval and press the SET key. Set the purge interval and
the unit (hour/day) by using the arrow keys. The interval must be
set between 1 hour...10 days. Press the OK key.
- Select Start-up purge and press the ON/OFF key.
- Start manual purge by selecting Manual purge and pressing the
START key.
4. Press the EXIT key to return to the basic display.
0507-012
Figure 47 Performing Chemical Purge
Using Serial Line
PURGE
Enter the PURGE command to start chemical purge immediately.
>purge
Purge started, press any key to abort.
>
The prompt '>' appears when the heating period is over. However, the
transmitter outputs are locked to the values measured before
performing chemical purge until the settling time is over.