User's Guide ______________________________________________________________________
92 __________________________________________________________________ M210796EN-A
where
xxx = Data recording quantity. For more information on available
quantities and units, see Table 2 on page 15. For more
information on optional quantities, see Table 3 on page 15 .
Example:
>dsel rh t tdf
RH T Tdf
>
Enter the command without parameters and press ENTER to display
current recording parameters.
View Recorded Data
If the device is provided with the optional display, the graphical
display shows the data of the selected quantities, one at a time. For
details about graphical display, see section Graphic History on page
58.
You may also dump the logged data to the serial line in numeric form
with the following commands.
DIR
Use the serial line and enter the DIR command to check the available
files.
The device records six files (six observation periods) for each selected
quantity. Thus, total amount of the files depends on the amount of the
selected quantities being at minimum 6 and at maximum 18. See
Table 10 on page 58.
Select, for example, two quantities (P and T). The last two columns
illustrate software information that is not essential for the user.
Example:
>dir
0 P latest 20 minutes 00-01-01 01:55:29 135 020A
1 P latest 3 hours 99-12-31 22:55:29 135 025A
2 P latest 1 day 99-12-30 23:17:59 135 040C
3 P latest 10 days 99-12-20 20:17:59 135 0802
4 P latest 2 months 99-10-25 14:17:59 135 080C
5 P latest 1 year 98-11-22 02:17:59 135 1003
6 T latest 20 minutes 00-01-01 01:55:29 135 020A
7 T latest 3 hours 99-12-31 22:55:29 135 025A