Vaisala HMP60 User's Guide Page 49

  • Download
  • Add to my manuals
  • Print
  • Page
    / 73
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 48
Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 47
Other Commands
Set Analog Output Mode
Use the AMODE command to show or set the operation mode of the
analog output. This command is not available on the HMP110D.
AMODE [ch1] [ch2]<cr>
where
ch1
=
Analog output mode for channel 1, range 0 ...3.
The options
are:
0 (
0 ... 1 V)
1 (0 .
.. 2.5 V)
2 (0 ... 5
V)
3 (1 ... 5
V)
ch2
=
Analog output mode for channel 2, range 0 ...3.
The options are the same as for channel 1.
Example (show current output modes):
amode
Ch1 output : 0 ... 1 V
Ch2 output : 0 ... 1 V
Example (set channel 1 to 0 ... 1 V and channel 2 to 0 ... 5 V):
amode 0 2
Ch1 output : 0 ... 1 V
Ch2 output : 0 ... 5 V