Chapter 9 _____________________________________________________________ Other functions
VAISALA _______________________________________________________________________ 31
CHAPTER 9
OTHER FUNCTIONS
Setting the alarm levels
MM70 is alarming by beeping and blinking the backlight. The alarm turns
on when the measured value is not between the alarm limits (permissable
area). Alarm level(s) can be set for only one quantity at a time. Set the
alarm levels as follows:
1. Open the MENU: press OPEN.
2. Select Functions, press .
3. Select Alarm, press .
4. Select Quantity and press SET to choose a quantity.
5. Select quantity, press
SELECT.(Only one quantity can be chosen at
a time. Only active quantities can be selected. To change the active
quantities, go to
MENU Display Quantities and units).
6. Select the first limit, press
SET (if the alarm function is ON, take it
OFF). Set the alarm level by using arrow buttons. Press +/- button
to choose the sign of the alarm level. Press OK to save the setting.
7. Select the second limit, then follow instructions in item 6. The alarm
will go off when the upper limit is exceeded or when lower limit is
gone below.
If you have only one limit, e.g. if you want to detect if the a
w
reaches
0.9, set the upper limit to 0.9 and the lower limit so low that it never
will be reached (for example 0). Thus, the MM70 will alert when RH
rises above 0.9 a
w
.