User's Guide _______________________________________________________________________
46 ____________________________________________________________________M210498en-A
2-point adjustment
Follow the steps from 1 to 4 and continue as follows:
5. Select 2-point adjustment. Press SELECT to continue.
6. Set the probe to a lower reference temperature. You can follow the
stabilization from the GRAPH display.
7. Press READY when the reading is stabilized. Give the lower
reference temperature by using the arrow buttons. Press OK.
8. Set the probe to the higher reference temperature. You can follow
the stabilization from the GRAPH display. Press READY when the
reading is stabilized.
9. Give the higher reference temperature by using the arrow buttons.
Press OK.
10. Confirm the adjustment, press YES. (By pressing NO you return
to adjustment mode display and no changes are made). If the
difference between two references is less than 30 °C, adjustment can
not be done.
11. Calibration is carried out. Press BACK to exit the adjustment mode
and EXIT to return to the basic display.
Last adjustment date
To see the last adjustment date, select Last adjustment date. You can check
this date also from the
Device information, see page 26.