Chapter 12 ________________________________ Calibration and adjustment of MM70 series probes
VAISALA _______________________________________________________________________ 43
When using Vaisala humidity calibrator HMK15, use the adapter fitting
(13.5 mm hole). Before starting the adjustment mode, take out the screw
covering the calibration button.
Relative humidity adjustment
LiCl-NaCl automatic adjustment
LiCl-NaCl automatic adjustment is a two-point adjustment in reference
humidities of 11.3 % (LiCl) and 75 % (NaCl). You don't have to feed the
reference values, the
MM70 displays the accurate value based on the
measured temperature and the Greenspan table stored into the memory of
the
MM70. Make the adjustment as instructed in the following section 2-
point relative humidity adjustment (select LiCl--NaCl autom. on item 8 and
follow the display instructions).
2-point adjustment
Low humidity references 0 % (Nitrogen)/ 11.3 % (LiCl) and higher
humidity references 75 % (NaCl)/ 97 % (K
2
SO
4
) are recommended for
two-point adjustment. Ensure that the difference between the humidity
references is more than 50 %.
Calibration button
(under the screw)