Chapter 10 _______________________________________Calibration and adjustment of transmitters
VAISALA _______________________________________________________________________ 39
6. Check that the probes are located in equal conditions and wait until
the readings are stabilized (can take 30 minutes or more). If you are
near the probes, do not breathe towards them.
7. Press ADJUST to continue adjusting.
8. Press To same as.., press SELECT. (MI70 recognizes always the port
to which the MMP70-series probe is connected).
9. Confirm by pressing
YES.
10. Adjustment is done. Press BACK and EXIT to return to the basic
display.
11. Switch off the MI70 and detach the calibration cable.
1-point adjustment by using a
calibrator
When adjusting a transmitter only in one reference condition, please take
care that the reference condition represents well the measuring
environment.
If you use humidity calibrator HMK15, please use the adapter fitting (13.5
mm) on the measurement hole.
Follow the steps from 1 to 4 and continue as follows:
5. Select the quantity press SELECT.
6. Remove the filter from the transmitter's probe and insert the probe
head into the reference condition.
7. Press
ADJUST to continue adjusting.
8. Select 1-point adjustment, press SELECT.
9. Press READY when the reading is stabilized in the reference
condition (takes about 30 minutes or more)
. You can follow the
stabilization from the GRAPH display.
10. Enter correct reference value by using the arrow buttons. Press OK.
11. Confirm by pressing YES.
12. Adjustment is done. Press BACK and EXIT to return to the basic
display.
13. Detach the calibration cable.