Operation modes of the GMP343
The GMP343 can operate in three different modes:
- RUN mode
- STOP mode
- POLL mode
In the RUN mode, the GMP343 sends measurement results at regular
intervals. The included parameters and the interval can be set by the user.
While in the RUN mode, the device will not answer any other command
except the S command, which stops the RUN mode and switches the
device to STOP mode.
The STOP mode is the default mode of operation for the GMP343, where
most of the commands listed in this document are available. You can enter
the STOP mode from the RUN mode with the S command, or from the
POLL mode with the OPEN<addr> command, where <addr> is the
unique address of the GMP343. While in the STOP mode, the GMP343
will respond to all unknown commands with the following string:
Unknown command.[cr][lf]>
The POLL mode is dedicated for bus interfacing. To work in the poll
mode, each device must have its own unique address. You can enter the
POLL mode by giving the CLOSE command in the STOP mode.
Only a single device can be accessed at a time in the POLL mode. In the
POLL mode, you can request a single measumement message from the
GMP343 using the SEND<addr> command. While in the POLL mode,
the GMP343 will ignore all unknown commands.