Setting the Averaging Filter
Use the averaging filter to reduce the noise of the measurement. Notice the
effect on the response time.
AVERAGE x <cr>
SAVE <cr>
x=0...60 seconds (default =10 s)
>average 20
AVERAGING (s) : 20
>save
EEPROM saved successfully (829 ms).
>
Setting the Smoothing Filter
Use the smoothing filter to reduce the noise of the measurement for long
averaging periods. Notice the effect on the response time.
SMOOTH x <cr>
SAVE <cr>
x=0...255 (default =0)
>smooth 10
SMOOTH : 10
>save
EEPROM saved successfully (829 ms).
>
Flushing the Filters (Resynchronization)
With this command you can flush the filters to get rid of the effect of the
past measurements. This feature is useful if long filtering lengths are used.
RESYNC <cr>
>resync
>