Serial Communication Settings
Save the changed settings by entering the SAVE command. The settings
are not valid until the device is reset.
SERI <cr>
SAVE <cr>
Baud rate: (300, 600, 1200, 2400, 4800, 9600, 19200, 38400, 57600,
115200)
Note: When baud rate is > 19200 the character spacing must be > 1 ms. If
the character spacing is not in use the device may lose some characters.
Data bits: (7/8)
Parity: (none, even, odd)
Stop bits: (1/2)
Key in the settings after the question mark and press ENTER.
>seri
BAUD RATE : 19200 ? 19200
DATA BITS : 8 ? 8
PARITY : NONE ? none
STOP BITS : 1 ? 1
>save
EEPROM saved successfully.
>