Vaisala HMP230 User Manual Page 154

  • Download
  • Add to my manuals
  • Print
  • Page
    / 163
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 153
HMP230 SERIES
Appendix 9: Re-gaining M210225en-B
148
4. DISPLAY/KEYPAD COMMANDS
Press the CL key to enter the command mode. Select More from the first
menu using the arrow keys and press ENT key. Select again More from the
second menu and press ENT. The following menu is displayed:
Select Regen and press ENT.
To change or inspect the re-gaining parameter settings select Regen. settings
and press ENT. To start re-gaining manually select Regeneration and press
ENT.
4.1. Re-gaining parameters
The re-gaining parameters are displayed in a sequence of
status of the automatic re-gaining
re-gaining interval
duration of heating
temperature difference
settling time
k parameter
Each setting can be changed by entering a new parameter value using the
arrow keys. NOTE: disconnect the security lock jumper for changing these
parameters! After the parameter sequence the menu Regen.settings/
Regeneration is displayed again. If you want to leave the parameters for the
main menu, press CL once or several times. This leaves the rest of the
parameters unchanged.
NOTE
To activate the new parameter settings immediately, the
transmitter must be switched off/on. Otherwise the
transmitter waits until the end of the old re-gaining
interval and then starts re-gaining using the new
parameters.
Page view 153
1 2 ... 149 150 151 152 153 154 155 156 157 158 159 ... 162 163

Comments to this Manuals

No comments