HMP230 SERIES
User's Guide M210225en-B
34
6. COMMANDS
The HMP230 transmitters use microprocessors; therefore their configuration
can be set according to the user's needs. This is done through commands,
either utilizing the menus on the local display or giving commands through the
serial interface (see Appendix 1). Most often the commands are used to
change the settings of the two analogue channels.
A limited range of commands can be given by using the three press switches -
up, down and enter - inside the transmitter housing. Four LEDs indicate the
command given with the up and down switches. The switches and LEDs are in
all HMP230 transmitters. LED commands can be used to calibrate the trans-
mitters (both humidity and temperature) or to calibrate the analogue outputs.
A full range of commands can be given through the display/keypad or through
the RS 232C serial bus. The commands can be used e.g. to select and scale the
outputs, to calibrate the humidity and temperature channels as well as the
analogue outputs and to set the serial interface.
6.1. Commands and security lock jumper
In order to prevent any tampering with the transmitter settings, the transmitters
cannot be calibrated, the analogue outputs set or the analogue output
quantities selected or scaled unless the security lock jumper has been discon-
nected. The commands involved are:
all LED commands
display/keypad commands:
Cali ð RH cal T cal
Analog outputs
Mode ð Analog outputs ð Mode
Scale
More ð More ð Frost
More ð Filt
serial commands:
CRH, CT, CTA, FCRH, ACAL; AMODE, ASEL, ASCL, FROST,
FILT
In the following, the description of these functions is preceded with a re-
minder of the security lock jumper:
Disconnect the security lock jumper!