8. MAINTENANCE
8.1. Reference measurements
Reference measurements are needed to verify whether the transmitter readings
are within specifications. This way the user can check if the transmitter needs
calibration or service.
The reference meter should preferably use the same technology as the instru-
ment checked, i.e. transmitters with capacitive sensors should be checked with
instruments using capacitive sensors. This minimizes the risk of errors caused
by different measurement techniques.
Whatever technique is used, make sure the reference instrument is at the same
temperature as the checked instrument in order to avoid errors caused by tem-
perature differences. The reference measurement should be made as close to
the checked sensor as possible and the readings should be read at the same
time, when possible.
The best reference measurements are made in laboratories. If it is possible to
take the transmitter out of process or control system, make the reference
measurement in a laboratory where the conditions are stable.
8.2. Self-diagnostics
The HMP230 transmitters go through a self-diagnostics procedure when the
power is switched on. If the procedure does not reveal any errors or faults, the
transmitter starts operating normally. If errors or faults are found, check first
whether the humidity and temperature sensors are damaged. If they are intact,
send the transmitter to Vaisala or a Vaisala representative for repairs. The
error messages the transmitter outputs are listed in Appendix 5.
If any errors occur during operation, the error messages are output on the local
display if the transmitter displays measurements; if the menus are used, error
messages are not output. The LEDs indicate errors at all times. During opera-
tion, however, the error messages are not output automatically through the se-
rial interface. If there is any reason to doubt that there is something wrong
with the transmitter, use command ERRS:
ERRS <cr>
If there are no error messages, only a prompt is displayed:
>ERRS <cr>
>