Chapter 7 ___________________________________________________ Calibration and Adjustment
VAISALA _______________________________________________________________________ 157
4. Press GRAPH to follow the RH and T stabilization on graphical
display:
0706-020
Figure 83 Following the RH Trend on Graphical Display
5. Follow the instructions on the display. Use the arrow buttons to
enter the actual humidity of the reference used.
0706-021
Figure 84 Ending Adjustment of Point 1
6. Continue adjustment of point 2 as instructed by the display:
0706-022
Figure 85 Proceeding to Adjustment of Point 2
7. Follow the purge progress and wait for stabilization of Point 2 as in
Figure 82 on page 156. Press READY when stabilized.
8. Use the arrow buttons to enter the actual humidity of the reference
used. Follow the instructions on the display to complete the RH
adjustment.
0706-023
Figure 86 Completing Adjustment of Point 2