102
GMT220 Series Carbon Dioxide Transmitters
for Industrial Applications
The Vaisala CARBOCAP
®
Carbon
Dioxide Transmitter Series GMT220 is
designed to measure carbon dioxide
in harsh and humid environments.
The housing is dust- and waterproof
to IP65 standards.
The GMT220 series transmitters
incorporate the advanced Vaisala
CARBOCAP
®
Sensor. The patented
sensor has unique reference
measurement capabilities. Its critical
parts are made of silicon; this gives
the sensor outstanding stability
over both time and temperature. By
lengthening the calibration intervals,
the user saves both time and money.
Interchangeable Probes
The user has a choice of
measurement ranges up to 20%
of CO
2
. The GMT221 is for higher
concentrations of CO
2
and the
The GMT220 transmitters withstand harsh and humid environments.
Features/Benefi ts
Incorporates Vaisala
CARBOCAP® - the silicon-
based NDIR sensor
IP65 protected against dust
and spray water
Several measurement ranges
Easy installation
Standard analog outputs
and two confi gurable relays
available
Applications include:
Horticulture and fruit storage
Greenhouses and mushroom
farming
Safety alarming and leakage
monitoring
Demand controlled ventilation
in harsh environments
GMT222 for lower concentrations
of CO
2
. The GMT220 probes are
interchangeable. They can be
removed and reattached or replaced
at any time – without the need for
calibration and adjustment. The
probes can be attached directly to
the transmitter body or, when used
with a cable, installed remotely into
hard-to-reach places or areas with
dangerously high levels of CO
2
.
The interchangeability of the GMT220
transmitter’s probes truly facilitates
eld maintenance.
The end user can carry out  eld
maintenance without any additional
equipment or heavy and expensive
calibration gas bottles by simply
replacing a probe.
Probes that have been replaced can
be sent to Vaisala for recalibration.
4-Carbon-Dioxide-Pages-100-119-Vaisala-Catalog-2011.indd 102 24.10.2011 9.21
103
Technical Data
Ref. B210827EN-C ©Vaisala 2011
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifi cations — technical included — are subject
to change without notice.
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
Performance
Measurement Ranges
GMT221 0 ... 2 %
for high concentrations 0 ... 3 %
0 ... 5 %
0 ... 10 %
0 ... 20 %
GMT222 0 ... 2000 ppm
for low concentrations 0 ... 3000 ppm
0 ... 5000 ppm
0 ... 7000 ppm
0 ... 10 000 ppm
Accuracy (including repeatability, non-linearity and
calibration uncertainty) at 25 °C and 1013 hPa
GMT221 ±(1.5 % of range + 2 % of reading)
(applies for concentrations above 2 % of full scale)
GMT222 ±(1.5 % of range + 2 % of reading)
Temperature dependence, typical -0.3 % of reading / °C
Pressure dependence, typical +0.15 % of reading/hPa
Long-term stability <±5 %FS/2 years
Response time (63 %)
GMT221 20 seconds
GMT222 30 seconds
Warm-up time 30 seconds, 15 minutes full specifi cations
Mechanics
Housing material
transmitter body ABS plastic
probe PC plastic
Housing classifi cation IP65
Weight:
GMT221 max. 280 g
GMT222 max. 300 g
Probe cable length 2 m and 10 m (optional)
Inputs and Outputs
Outputs 0 ... 20 or 4 ... 20 mA
and 0 ... 10 V
Resolution of analog outputs 12 bits
Recommended external load:
current output max. 400 Ohm
voltage output min. 1 kOhm
Two pre-or user-defi ned relay
outputs
Relay contacts max. 30VAC/60VDC, 0.5A
Connections screw terminals, 0.5 ... 1.5 mm
2
Operating voltage nominal 24 VAC/DC
Power consumption <4 W
Operating Environment
Operating temperature -20 ... +60 °C (-4 ... +140 °F)
with display 0 ... +50 °C (+32 ... +122 °F)
Storage temperature -30 ... +70 °C (-22 ... +158 °F)
Operating pressure (compensated range) 700 ... 1300 hPa
Humidity 0 ... 100 %RH, non-condensing
Electromagnetic compatibility EN61326-1, Generic Environment
Accessories
Spare probe GMP221, GMP222
(use the order form to defi ne measurement range etc.)
Clips (2 pcs) for attaching the probe 25245GM
Mounting fl ange for the probe GM45156
Probe cables
2 m 25665GMSP
10 m 210848GMSP
Calibrator for interchangeable probes GMK220
Wall Assembly Plate GM45160
In-soil adapter for probe 211921GM
Serial COM adapter 19040GM
Calibration adapter for probe 26150GM
GMT220 transmitter GMP222 probe
GMP221 probe
Dimensions
Dimensions in mm (inches)
119 (4.69)
31.8 (1.25)
119 (4.69)
109 (4.29)
18.5 (0.73)
155 (6.102)
18.5 (0.73)
4-Carbon-Dioxide-Pages-100-119-Vaisala-Catalog-2011.indd 103 24.10.2011 9.21
Page view 104
1 2 ... 100 101 102 103 104 105 106 107 108 109 110 ... 161 162

Comments to this Manuals

No comments