Technical Data
Ref. B210765EN-C ©Vaisala 2011
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
Dew point temperature (°C)
0
20 40 60
0
40
-60
-80
Process temperature (°C)
10
20
30
50
60
70
-10
-20
-30
-50
-40
-70
3010
50
Accuracy < ± 2°C Td
Dew point measurement range lower limit
Water vapor saturation line
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
Dew Point Temperature
Measurement range (typical) -60 ... +60 °C (-76 ... +140 °F)
Analog output scalings
Option A -80 ...+20 °C (-112 ... +68 °F) T
d
Option B -60 ... +60 °C (-76 ... +140 °F) T
d
Option X free scaling
(when the dew point is below 0 °C (32 °F),
the transmitter outputs frost point)
Accuracy with DRYCAP
®
180M ±2 °C (±3.6 °F)
(see graph below)
Operating Environment
Temperature 0 ... +60 °C (32 ... +140 °F)
higher temperature peaks Short-term OK
Relative humidity 0 ... 100 %RH
Pressure 0 ... 20 bara (0 ... 290 psia)
Sample flow rate no effect
Output
Analog output 4 ...20 mA
Resolution for analog output ±0.002 mA
Typical temperature dependence 0.0008 mA/ °C
Serial line for service use RS232
General
Sensor DRYCAP
®
180M
Optimal sensor for refrigeration dryers DRYCAP
®
180S
Operating voltage 18 - 35 VDC, 20 - 28 VAC
Power consumption at 24 VDC max. 220 mA
External load for analog output max. 500 Ω
Optional connection cable with
DMT242 connector 2 m or 10 m
Connector for supply voltage and
signal output
max. wire size 0.75 mm
2
max. cable diameter 6.5 mm /PG7
Service cable for serial interface
RS232 product code DMT242RS
Probe material (wetted parts) stainless steel
(AISI 316L)
Sensor protection stainless steel
sintered filter (HM47280)
Mechanical connection G½” ISO228-1 thread with
bonded seal ring (U-seal)
Eletronics housing material plastic (ABS/PC)
Housing classification IP65 (NEMA4)
Storage temperature range -40 ... +70 °C, (-40 ... +158
°F)
Complies with the EMC standard EN61326-1, Electrical equipment
for measurement, control and laboratory use - EMC requirements;
Industrial environment.
DEW POINT ACCURACY VS. MEASUREMENT CONDITIONS
Response time 63 % [90 %] at +20 °C gas temperature
Flow rate >1 l/min and 1 bar pressure
-60 -> -20 °C Td (-76 -> -4 °F Td) 5 s [10 s]
-20 -> -60 °C Td (-4 -> -76 °F Td) 45 s [10 min]
Accuracy with DRYCAP
®
180S ±2 °C (±3.6 °F)
(see graph below)