Technical Data
Performance
Dew point accuracy ±2 °C (±3.6 °F)
Pressure accuracy at 23 °C (73.4 °F) ±0.4 %FS
Pressure temperature dependence ±0.1 %FS/10 °C (18 °F)
Temperature accuracy
0 ... 40 °C (+32 ... +104 °F) ±0.5 °C (± 0.9 °F)
-40...80 °C (-40 ... +176 °F) ±1 °C (± 1.8 °F)
PPM accuracy (7 bar) ±(14 ppm + 12% of reading)
Sensor response time:
Pressure response time < 1 s
Dew point response time 63% [90%] at 20°C and 1 bar
-50 -> -10 °C Tdf 5 s [ 10 s]
-10 -> -50 °C Tdf 10 s [ 2.5 min]
Operating Environment
Operating temperature of electronics -40 … +60 °C
(-40 ... +140 °F)
Operating Pressure 0 … 50 bar (0…725 psi)
Outputs
Analog Outputs (2 channels)
current output 0 … 20 mA, 4 … 20 mA
voltage output 0 … 5V, 0 …10V
Accuracy of analog outputs ± 0.01 V / ± 0.01 mA
Digital output RS-485, non-isolated, Vaisala protocol
Connector 4-pin M8
General
Sensor Vaisala MPS1 multiparameter sensor
Operating voltage 21 ... 28 VDC, current output
20 ... 28 VDC, voltage output and/or use
in cold temperatures (-40 ... -20 °C (-40 ... -4 °F))
15 ... 28 VDC, RS485 only
Supply current
during normal measurement 20 mA + load current
during self-diagnostics 300 mA + load current
External load for
current output max. 500 Ohm
voltage output min. 10 kOhm
Housing material AISI316L
Housing classification IP65 (NEMA4)
Sensor protection Mesh filter AISI303, grade 18 µm
Storage temperature range
transmitter only -40 … +80 °C (-40 ... +176 °F)
shipment package -20 … +80 °C (-4 ... +176 °F)
Mechanical connection ISO G1/2”, NPT 1/2", UNF 3/4"-16
Recommended calibration interval 2 years
Weight (ISO1/2") 190 g (6.70 oz)
Complies with EMC standard EN61326-1, Electrical equipment for
measurement, control and laboratory use - EMC requirements;
Industrial environment
Ref. B211159EN-B ©Vaisala 2011
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
DPT146 Dewpoint Measurement Accuracy
Calculated Parameters
ppm moisture, by volume 1 ... 40 000 ppm
Dew point, converted to
atmospheric pressure -75 … +30 °C (-103 ... +86 °F)
Measured Parameters
Dew point -60 ... +30 °C (-76 ... +86 °F)
Pressure, absolute 1 ... 10 bar (14.5 ... 145 psi)
Temperature (available if output
RS-485 only selected) -40 ... +80 °C (-40 ... +176 °F)
Accessories
Connection cable for MI70 indicator /DM70 meter 219980
USB connection cable 219690
Sampling cells DMT242SC, DMT242SC2, DSC74, DSC74B,
DSC74C
Flange DM240FA
Loop-powered external display 226476
ISO 1/2" plug 218773
NPT 1/2" plug 222507
Relative humidity 0...100 %
Measured gases Air/ non-corrosive gases
Sample flow rate no effect on measurement accuracy
0 20
20
0
10
30
-10
-20
-30
10
30 40 50 60
40
Dewpoint temperature (°C)
Temperature of measured gas (°C)
-40 -30 -20 -10 70 80
-40
-50
-60
Not recommended for
continuous measurement
Dewpoint measurement low limit
-70
Accuracy not specified
Accuracy
±2 °C Td
-80