DMT142 Miniature Dewpoint Transmitter
for OEM Applications
The Vaisala DRYCAP
®
Dewpoint Transmitter DMT142 is an
ideal choice for small compressed air dryers, plastic dryers
and other OEM applications.
Features/Benefits
Miniature size dew point transmitter for e.g. small
industrial dryer applications
Vaisala DRYCAP® technology with auto-calibration
Calibration interval of two years
Dew point measurement range -60 ... +60 °C
(-76 ... +140 °F) with an accuracy ±3 °C (±5.4°F)
Withstands condensation
Fast response time
Compatible with Vaisala DRYCAP® Hand-Held
Dewpoint Meter DM70
NIST traceable (certificate included)
Vaisala DRYCAP®
The Vaisala DRYCAP
®
Dewpoint Transmitter DMT142 is
a miniature dew point measurement instrument. The
transmitter can be installed directly into pressurized
systems at 50 bar (725 psia) maximum pressure. The long-
term high performance is achieved with Vaisala DRYCAP
®
technology.
The sensor fully withstands getting wet, and therefore, the
transmitter performs exceptionally well in applications
that occasionally experience process water spikes, such as
pipeline condensation during a system failure or start-up.
The sensor is also immune to particulate contamination,
oil vapor and most chemicals, and is insensitive to the
flow rate.
Long Calibration Interval
The calibration interval of the DMT142 is two years.
Additionally, the Vaisala DRYCAP
®
Hand-Held Dewpoint
Meter DM70 can be used to confirm the performance of the
DMT142 without disconnecting the transmitter.
For any adjustment needs, the transmitter can be sent to
Vaisala Service.
The auto-calibration software works on-line while the
process is running. If the measurement accuracy is not
confirmed, corrections are made automatically.
Technical Data
Measured Variables
DEW POINT TEMPERATURE
Measurement range(typical)
Standard version -40 ... +60 °C (-40 ... +140 °F) T
d
Extended version -60 ... +60 °C (-76 ... +140 °F) T
d
SF
6
version -50 ... +60 °C (-58 ... +140 °F) T
d
Analog output scalings
option 1 -80 ...+20 °C (-112 ... +68 °F) T
d
option 2 -60 ... +40 °C (-76 ... +104 °F) T
d
option 3 free scaling
Accuracy ±3 °C (±5.4 F) T
d
(see graph below)
(when the dew point is below 0 °C (32 °F),
the transmitter outputs frostpoint)
Easy Installation
The DMT142 has a variety of features to choose from,
including different output and installation options. Due to
its small size and light weight, the DMT142 is quickly and
easily installed in tight spaces or in small-size pipelines.