Technical Data
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
Ref. B210901EN-B ©Vaisala 2011
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
Relative Humidity
Measurement range 0 ... 100 %RH
Accuracy (including non-linearity,
hysteresis and repeatability)
+2 %RH (0 ... 90 %RH)
+3 %RH (90 ... 100 %RH)
Response time (90 % at +20 °C/+68 °F)
in still air (with sintered filter) 60 s
Typical temperature dependence
of electronics 0.02 %RH/°C (0.02 %RH/°F)
Humidity sensor HUMICAP
®
180R
Power Supply
Operating voltage 10 ... 35 VDC
In modules with analog outputs the supply range depends to a
certain extent on the selected output range.
Current consumption without sensor head
warming or re-gaining option
HMM211 & HMM213 12 mA at 35 VDC
Temperature
Measurement range -70 ... +180 °C (-94 ... +356 °F)
Typical accuracy of electronics at +20 °C
+0.1 °C (+0.18 °F)
Typical temperature dependence of electronics 0.0025°C/°C
(0.0025 °F/°F)
Temperature sensor in RH+T probe:
HMM211 and HMM213 Pt 100 RTD
IEC 751 1/3 Class B
HMM212 Pt 1000 RTD
IEC 751 1/3 Class B
Additional temperature probe Pt 100 RTD
IEC 751 1/4 Class B
Configuration Options
Compatible with modules:
PROBES HMM211 HMM212 HMM213
RH+T yes yes yes
Dewpoint (heated
composite sensor) yes yes
Temperature yes yes
CABLE LENGTHS
RH+T and Dewpoint probes 65, 150 and 300 cm
Temperature probe 150 and 300 cm
CHEMICAL PURGE
Automatically takes place at power-up
HUMICAP
®
is a registered trademark of Vaisala.
General
Operating temperature range
Probe -70 ... +180 °C (-94 ... +356 °F)
Electronics -5 ... +55 °C (+23 ... +131 °F)
Storage temperature range
(Electronics) -40 ... +70 °C (-40 ... +158 °F)
Sensor protection (standard) stainless steel sintered filter
Connections screw terminals for 0.5 ... 1.5 mm
2
wires
Outputs
Two analog outputs selectable
HMM211 0 ... 1 V, 0 ... 5 V, 0 ... 10 V
0 ... 20 mA
HMM212 4 ... 20mA (loop powered)
Digital output
HMM213 RS232
Dimensions
Dimensions in mm (inches)
60 (2.36)
63 (2.48)
ø6
ø12
90 (3.54)
~92 (3.62)
(0.24)
(0.47)
26 (1.02)
M10×1.0
HMM211
60(2.36)
49(1.93)
63(2.48)
57(2.24)
ø12
90(3.54)
ø3.5
30(1.18)
36(1.42)
(0.47)
(0.14)
26(1.02)
M10×1.0
HMM212
~65(2.56)
(0.14)
26 (1.02)