Technical Data
Ref. B211160EN-A ©Vaisala 2011
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
Accessories
Connection cable for the MI70/DM70 hand-held 219980
USB connection cable 219690
Protection plug for connector 218675SP
1.5 m Shielded PUR cable with 90° connector 231519SP
3m Shielded PUR cable with 90° connector 231520SP
5 m Shielded PUR cable with 90° connector 231521SP
10 m Shielded PUR cable with 90° connector 231522SP
3.0m Shielded FEP cable with straight connector 226902SP
Weather shield ASM210326SP
Calculated Parameters
Pressure, normalized to 20 °C (68 °F) 1 ... 12 bar (14.5…174 psi)
SF6 or SF6/N2 mixture density 0 ... 100 kg/m
3
ppm moisture, by volume 40 ... 40 000 ppm
Dewpoint, converted to atmospheric
pressure -65 … +30 °C (-85 ... +86 °F)
Measured Parameters
Dewpoint -50 ... +30 °C (-58 ... +86 °F)
Pressure, absolute 1 ... 10 bar (14.5 ... 145 psi)
Temperature -40 ... +80 °C (-40 ... +176 °F)
Operating Environment
Operating temperature of electronics -40 … +60 °C (-40 ... +140 °F)
Operating Pressure 0 … 50 bar (0…725 psi)
Relative humidity 0...100 %
Measured gases SF
6
, SF
6
/N
2
mixture
General
Sensor Vaisala MPS1 multiparameter sensor
Operating voltage 15 ... 28 VDC
20 ... 28 VDC in cold temperatures (-40 ... -20 °C (-40 ... -4 °F))
Supply current, during normal measurement 20 mA
during self-diagnostics max. 300 mA pulsed
Housing material AISI316L
Housing classification IP65 (NEMA4)
Weather shield to be used for continuous outdoor installations
Storage temperature range
transmitter only -40 … +80 °C (-40 ... +176 °F)
shipment package -20 … +80 °C (-4 ... +176 °F)
Mechanical connection DILO DN20, ABB Malmkvist, or
Alstom G1/2” compatible connector
Every connection is helium leak tested at the factory.
Dimensional drawings See the document B211165EN-A
Weight (with DILO adapter) 765 g (27.0 oz)
Complies with EMC standard EN61326-1, Electrical equipment for
measurement, control and laboratory use - EMC requirements;
Industrial environment, Tested levels
EN/IEC 61000-4-2, Electrostatic Discharge 8kV con / 15kV air
EN/IEC 61000-4-3, RF field immunity 10V/m (80MHz-4.2GHz)
EN/IEC 61000-4-4, Electric Fast Transient ±2kV power and signal
EN/IEC 61000-4-5, Surge ±2kV power line to ground / ±1kV signal
line to ground and power line to line
EN/IEC 61000-4-6, Conducted RF
Immunity
10Vemf power line
and digital output
Mechanical vibration
EN/IEC 60068-2-6, Fc Sinusoidial vibration ± 6 g, 5-500 Hz sweep
60 min/axis, 3-axis
Outputs
Digital output RS-485, non-isolated, Vaisala protocol
Connector 4-pin M8
Performance
Dewpoint accuracy ±3 °C (±5.4 °F), see graph below
Dewpoint stability typical drift < 2 °C (3.6 °F) /5 years
Pressure accuracy at 23 °C (73.4 °F) ±0.4 %FS
Pressure temperature dependence ±0.1 %FS/10 °C (18 °F)
Pressure stability typical drift < 1 %FS /5 years
Temperature accuracy
0 ... 40 °C (+32 ... +104 °F) ±0.5 °C (± 0.9 °F)
-40...80 °C (-40 ... +176 °F) ±1 °C (± 1.8 °F)
Density accuracy (pure SF6, 1 ... 10 bara)
0 ... 40 °C (+32 ... +104 °F) ±1 %FS
-40 ... +60 °C (-40 ... +140 °F) ±2.2 %FS
PPM accuracy, typical (5...1000 ppm, 7
bar)
±(7 ppm + 15% of reading)
Sensor response time:
Pressure response time < 1 s
Dewpoint response time* 63% [90%] at 20°C and 1 bar
-50 -> -10 °C Tdf 5 s [ 10 s]
-10 -> -50 °C Tdf 10 s [ 2.5 min]
* system equilibrium related response time is typically longer
0 20
20
0
10
30
-10
-20
-30
10
30 40 50 60
40
Dewpoint temperature (°C)
Temperature of measured gas (°C)
-40 -30 -20 -10 70 80
-40
-50
-60
Not recommended for
continuous measurement
Dewpoint measurement low limit
-70
Accuracy not specified
Accuracy
±3 °C Td
DPT145 Dewpoint Measurement Accuracy