USER'S GUIDE____________________________________________________________________
68 __________________________________________________________________ M211095EN-C
Example 1
- Output mode: voltage
- Offset: 0.0
- Gain: 0.1
With the above settings, the analog output voltage range is from 0 V (0
m/s) to 7.5 V (75 m/s). When the measured wind speed is 10 m/s, the
output voltage is 1.0 V. See the formula below:
Output = 0.0 + 10 × 0.1 = 1.0 V
Example 2
- Output mode: current
- Offset: 0.004
- Gain: 0.0002
With the above settings, the analog output current range is from 4 mA
(0 m/s) to 19 mA (75 m/s). When the measured wind speed is 10 m/s, the
output current is 6 mA. See the formula below:
Output = 0.004 + 10 × 0.0002 = 6.0 mA.
Limitations for Output Signals
You can specify the minimum and maximum values for analog output
with the configuration parameters. The output is fixed to the specified
values, and the unit depends on the selected analog output mode.
Example
To limit analog output 1 in voltage mode to a range of 0.1 … 5 V, set the
analog output minimum value to 0.1 and the analog output maximum
value to 5. Enter the following commands:
S aout1minv,0.1
S aout1maxv,5
For more information on the parameters, see Table 18 on page 82.