Example of the command and response:
W5
W5P1200013.2TDE
Interpretation of the example message:
-
02H (<STX>, start of transmission)
- Sensor ID: W
- Averaging in seconds: 5
- Sensor status: P = pass
- Wind direction: 120
- Wind speed: 0013
- . (dot character)
- Wind speed unit: T = meters per second
- Checksum (most significant digit): D
- Checksum (least significant digit): E
-
03H (<ETX>, end of transmission)
Missing Readings
If data is missing due to a measurement problem, the WS425 A/B
ASCII message reports 999.9 for wind speed.