Chapter 7 ____________________________________________________________ Troubleshooting
10. Press # on your keyboard and hold down the key for at least
five seconds. WMT700 responds by sending the following
string:
Ok
Restoring COM1 and COM2 settings…
Updating the settings will be completed in a few seconds.
When the operation is done, WMT700 sends the following
response:
Done. Rebooting…
WMT700 has now applied the settings and will reboot using
the new settings. The restored settings are shown in Table
57 below.
11. Before you start configuring new values to WMT700, you
need to apply the new settings to HyperTerminal. In the
Call menu, click Disconnect to close the connection.
12. On the File menu, click Properties.
13. In the Properties dialog, click Configure.
14. Select 9600 in the Bits per second combo box.
15. Close the Properties window.
16. On the Call menu, click Call.
Table 57 Restored Serial Port Settings
4 = 9600 bauds per second
Response delay in milliseconds
com1_protocol
com2_protocol
The serial port settings have now been restored to known values.
You can send the OPEN command and start configuring
WMT700. For information on the configuration commands, see
section Configuration on page 110.
NOTE
You can also use other terminal programs
, such as Tera Term.