Chapter 2 __________________________________________________________ Product Overview
VAISALA _______________________________________________________________________ 15
Constant Output Probe
The constant output probe is a testing accessory that can be used to check
transmitter's functions and measurement signal transfer chain all the way
to the control system. The constant output probe does not measure
humidity and temperature; instead, it outputs constant humidity and
temperature readings.
The values output by the constant output probe are specified when
ordering. These values are written on an additional label on the probe
body.
The procedure for using the constant output probe is simply to replace
the original probe for the duration of the testing:
1. Disconnect the normal probe from transmitter.
2. Connect the constant output probe to the transmitter.
3. Check that all used outputs (analog, display, serial line) show the
correct measurement values.
4. After checking the outputs, disconnect the constant output probe
and reconnect the original probe.
NOTE
The transmitter goes to the error state for a short time when the probe is
changed. This is normal.