Chapter 5 ______________________________________________________________ Maintenance
VAISALA _______________________________________________________________________ 69
Offset and Gain Adjustments
Offset and gain adjustments are performed in the same way (they differ
only in internal calculations), so this section is valid for both methods.
4. Connect the 25917ZZ HMI41-connection cable to the SERVICE PORT
connector on the HMT130 motherboard (see Figure 10 on page
24).
5. Connect the other end of the connection cable to the EXT connector
located on the bottom of the HMI41 indicator (see Figure 20 on
page 67).
6. Turn on both devices. Wait until the following display appears:
RH
7. Select the appropriate adjustment mode with the and buttons.
Select offset adjustment if the humidity is < 65 %RH and gain
adjustment if the humidity is 65 %RH:
RH
or
RH
Press ENTER to confirm the selection. A display similar to the
following appears:
RH %
NOTE
In case of using a HMT130 transmitter with display option the text
"HMI41 adjustment mode" is shown at the upper row of HMT130
display.
The next sections describe how to perform offset and gain adjustments in
two different ways.