Chapter 6 ___________________________________________________________ Troubleshooting
VAISALA _______________________________________________________________________ 75
CHAPTER 6
TROUBLESHOOTING
This chapter describes error messages and analog output error behavior,
introduces some common problems, their probable causes and remedies
and provides contact information for Technical Support.
Error Codes
HMT130 software includes number of different self diagnostics, such as
flash and program memory checksum, probe communication status,
probe checksum, operation voltage check, oscillator fault check etc.
At startup software checks factory/user settings checksum, program
memory checksum and oscillator fault status. Other checks are made
during runtime. Available error codes are shown in the table below.
Table 7 Error Codes and Texts
Code Description Error Text
1 Probe T measurement error Probe T meas
2 Probe RH measurement error Probe RA meas
3 Probe communication error Probe communication
4 Probe checksum error Probe checksum
5 Probe message form error Probe message form
6 Program's flash checksum error Program code checksum
7 Current settings checksum error
(RAM)
Settings checksum
8 Factory flash not initialised Factory defaults empty
9 User flash not initialised User defaults empty
10 Voltage is too low to operate
correctly
Voltage too low
11 Measurements not available Measurements not available
12 Oscillator fault bit active HW fault 1
13 Analog output quantity invalid Analog output quantity
invalid
14 Display quantity invalid Display quantity invalid
Error text is shown by using the ERRS command via the serial interface.
In case of using a HMT130 transmitter with display option the error code
is shown at the display in format "ERR:Code-1[.Code-2][.Code-n]".
Multiple errors are separated with a dot between the error codes.