NOTE
Select the size and type of the fastening screws according to the wall
material (wood, stone, etc.). Even though using all four screws is
strongly recommended, the HMT130 enclosure fastening holes are
initially covered with a thin plastic membrane, so less than four screws
could be also be used without sacrificing the ingress protection (IP) class
of the enclosure. The diameter of the fastening screws is typically
between 3.5 and 4mm.
Installation with Rain Shield
The installation kit with rain shield includes a metal mounting plate and
a rain shield for the transmitter. Vaisala order code: 215109.
1
2
31
2
3
1011-161
Figure 3 HMT130 Installation with Rain Shield
1. Fasten the metal mounting plate to the wall or pole with screws
(see Figure 3 above). Note the arrow on the mo
unting plate. Attach
the mounting plate with the arrow pointing upwards.
2. Drill holes for the screws in the HMT130 frame, and fasten
HMT130 Humidity and Temperature Transmitter to the metal
mounting plate with four (M4) screws.
3. Fasten the rain shield to the metal mounting plate with two (M6)
mounting screws.